Hold the sensor with two fingers on the base, and gently remove the introducer needle. Do NOT rotate
the introducer needle when removing. Dispose of the needle in sharps container.
11 12
13
14
ಘ
Wait five minutes after insertion before connecting the transmitter to the sensor to make sure the sensor
is ready to communicate. Make sure that the site is not bleeding before connection. If bleeding occurs,
apply pressure using a sterile gauze or clean cloth for three minutes.
a. When bleeding stops, attach the transmitter to the sensor.
CAUTION: If bleeding does NOT stop, do NOT connect the transmitter to the sensor.
b. If bleeding does not stop, do the following:
i Remove the sensor and discard.
ii Reapply pressure using a sterile gauze or a clean cloth until the bleeding stops.
iii Insert a new sensor in a different location.