The percentage of Guardian RT readings within 20 percent and 30 percent of YSI readings was analyzed
according to time after sensor insertion and according to the glucose-concentration range (as determined by
the YSI analyzer). See the following table.
Percentage of Guardian RT values within
20% of YSI laboratory readings
Percentage of Guardian RT values within
30% of YSI laboratory readings
Glucose Range
(mg/dl)
During first 60
hours of sensor
wear
After 60 hours of
sensor wear
During first 60
hours of sensor
wear
After 60 hours of
sensor wear
40-80
a
62-82% 39% 62-82% 39%
81-120 57-66% 48% 72-84% 66%
a. Agreement to within ± 1.1 mmol/L (20 mg/dL) for glucose readings 4.4 mmol/L (80 mg/dL)
Performance of the Guardian RT was evaluated according to the length of time since calibration. This data
is not conclusive because of the limited number of data points during the final 3 hours of the 12-hour
calibration cycle, i.e., 10. In contrast, 3-hour time bins, earlier in the 12-hour cycle, contained hundreds of
data points. This may suggest that calibrations are often required prior to the 12-hour calibration cycle.
Appendix A
218