Basal
Rate increments: 0.05 units/hour
Delivery: 0.05 - 35 units/hour (maximum units: 35/hour)
Factory maximum setting: 2.0 units/hour
3 patterns maximum, each with 48 rates
BG Target
Maximum targets: 8
Range: 60 - 250 mg/dL (3.3 - 13.9 mmol/L)
Warning limits: less than 90 or greater than 140 mg/dL (less than 5.0 or greater than 7.8 mmol/L)
Bolus delivery
Insulin delivered/stroke: 0.05 u
Fluid delivered/stroke: 0.5 μL (microliters)
Time between/delivery strokes: 2 seconds
Delivery rate (per minute): 1.5 u
Bolus history
Maximum records shown: 24
Bolus units
Increments: 0.1 units
Bolus Wizard feature
See the Bolus Wizard feature specifications section in this chapter for information.