Medtronic Paradigm Link Podręcznik Użytkownika Strona 81

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 80
12
To purge air bubbles that have risen to the top of the reservoir, push up on the plunger until you see
insulin in the tubing.
13
Without pulling, turn the plunger counter-clockwise to remove it from the reservoir.
11 12
13
Changing your infusion set
Removing the reservoir
Each time you remove and replace a reservoir in your pump, you have to rewind your pump, and fill the
infusion set with insulin.
1
Remove the entire infusion set from your body.
2
If attached, remove the activity guard.
3
Turn the tubing connector half-turn counter
clockwise, then pull the reservoir and connector
out from the pump.
tubing connector
half-turn counter-clockwise
4
Safely dispose of the used reservoir and infusion set item in a sharp's container.
5
You must now rewind your pump as described in the next section.
Starting on insulin
62
Przeglądanie stron 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 261 262

Komentarze do niniejszej Instrukcji

Brak uwag