pattern for very low transmitter battery, leave the transmitter on the charger for eight hours to recharge.
If the red light is still flashing after eight hours, call our 24 Hour HelpLine. It may be time to replace your
transmitter.
Question: I had my transmitter on the charger for a day. Will this damage my transmitter?
Answer: It will not damage the transmitter. You cannot overcharge it.
Question: What should I do if the transmitter’s green light did not flash after removing it from the
charger?
Answer: Reconnect the transmitter to the charger for at least one minute, remove it and watch the
transmitter’s green light flash and then turn off.
Question: What should I do if the transmitter’s green light does not flash when connected to the sensor?
Answer: Is the sensor inserted in the body? If it is not inserted, the transmitter will not flash green or send
signals to a monitoring device.
If the sensor is inserted in the body, you need to disconnect the transmitter from the sensor, wait at least
one minute and then reconnect. If the green light still does not flash, charge the transmitter.
Question: Why didn’t I see the transmitter’s green light flash after connecting it to the tester?
Answer: Check the connection. If you still do not see a green light flash, fully recharge the transmitter
battery. Test the transmitter with the tester. If you still do not see a green light flash, call the HelpLine. It
may be time to replace your transmitter.
Troubleshooting and alarms
174