2 Select Glucose Limits, then press ACT. The SET GLUCOSE LIMITS 1 screen appears. The Low Glucose
Limit flashes 80 mg/dL (4.4 mmol/L).
The start time for your first pair of Glucose Limits is midnight and cannot be changed.
3 Select your Low Glucose Limit. For the 523 and 723 pumps, the value must be between 40 and
390 mg/dL (2.2 and 21.6 mmol/L). To switch off the Low Glucose Limit, reduce it below 40 mg/dL
(2.2 mmol/L) to select OFF. For the 523K and 723K pumps, the value must be between 90 and
390 mg/dL (5.0 and 21.6 mmol/L). To switch off the Low Glucose Limit, reduce it below 90 mg/dL
(5.0 mmol/L) to select OFF.
4 Press ACT to select your High Glucose Limit. For the 523 and 723 pumps, the screen flashes 240 mg/dL
(13.3 mmol/L). The limit you select must be between 50 and 400 mg/dL (2.8 and 22.2 mmol/L). Your
High Glucose Limit must also be at least 10 mg/dL (0.6 mmol/L) above your Low Glucose Limit. To
switch off the High Glucose Limit, reduce the High Glucose Limit below 50 mg/dL (2.8 mmol/L) to
select OFF. For the 523K and 723K pumps, the screen flashes 280 mg/dL (15.5 mmol/L). The limit you
select must be between 100 and 400 mg/dL (5.6 and 22.2 mmol/L). Your High Glucose Limit must also
be at least 10 mg/dL (0.6 mmol/L) above your Low Glucose Limit. To switch off the High Glucose
Limit, reduce the High Glucose Limit below 100 mg/dL (5.6 mmol/L) to select OFF.
5 Press ACT. The SET START TIME 2 screen appears. The dashes under the screen name flash. The first
pair of Glucose Limits is now set.
If you do not need a second pair of Glucose Limits, press ESC and skip to the next section. If you want
to set up a second pair of Glucose Limits, follow steps 6 through 11 below.
6 To set a second pair of Glucose Limits, select the time of day you want this pair to become active.
7 Press ACT. The SET GLUCOSE LIMITS 2 screen appears. The default Low Glucose Limit flashes (OFF).
8 Select your Low Glucose Limit. For the 523 and 723 pumps, the value must be between 40 and
390 mg/dL (2.2 and 21.6 mmol/L). For the 523K and 723K pumps, the value must be between 90 and
390 mg/dL (5.0 and 21.6 mmol/L).
9 Press ACT. The default High Glucose Limit flashes (OFF).
10
Select your High Glucose Limit. For the 523 and 723 pumps, the value must be between 50 and
400 mg/dL (2.8 and 22.2 mmol/L). For the 523K and 723K pumps, the value must be between 100 and
400 mg/dL (5.6 and 22.2 mmol/L).
11 Press ACT. The SET START TIME 3 screen appears.
The second pair of Glucose Limits (Glucose Limits 2) is now set. If you do not need any more Glucose
Limits pairs, press ESC. If you want to set more Glucose Limits, then repeat steps 6 through 11 above
for each pair, up to a maximum of eight.
You are ready to set up your High Repeat.
Sensor features106