Example 2: Blood glucose above target (high blood glucose) and no active insulin
The next day, Michael wakes up before school. Before eating the same breakfast, he tests his blood
glucose with his meter and finds it to be 200 mg/dL (11.1 mmol/L), which is above his target of
120 mg/dL (6.6 mmol/L). His blood glucose reading is automatically sent to his pump.
When prompted by the Bolus Wizard feature, he enters his carbohydrate amount of 60 grams in the
ENTER FOOD screen. Based on his settings, the pump will suggest that he take 6.0 units of insulin.
Food estimate:
Carb grams ÷ Carb ratio = Units of insulin
60 g ÷ 15 g/u = 4 units
Correction estimate:
(Current BG - High BG Target) ÷ Insulin sensitivity = Units of insulin
(200 mg/dL - 120 mg/dL) ÷ 40 mg/dL/u = 2 units, or
(11.1 mmol/L - 6.6 mmol/L) ÷ 2.2 mmol/L/u = 2 units
Total bolus estimate:
Food estimate + Correction estimate = Units of insulin
4 units + 2 units = 6 units