NOTES:
If a Dual Wave bolus is less than the estimate due to the max bolus limit or a user change, the square
(sq) portion is reduced first.
Based on the Active Insulin Time setting you choose, the pump
keeps track of how much insulin is still active in your body. This is
shown as Active Ins. or Act. Ins. in the STATUS, SET BOLUS, SET
NORMAL BOLUS, SET SQUARE BOLUS, SET DUAL BOLUS TOTAL, and
SET EASY BOLUS screens. This prevents stacking of insulin, and
lowers the chances of hypoglycemia.
The active insulin amount is also shown in the Bolus Wizard’s ESTIMATE DETAILS screen but is
calculated differently than in the above screens. The active insulin amount calculated in the
ESTIMATE DETAILS screen includes the insulin that has already been delivered and insulin that is going
to be delivered by the active Square Bolus. To show this difference in calculation, the active insulin
appears with an asterisk (*Active Insulin) in this screen.
Pump specifications188