Use your reservoir immediately after you fill it. Do not store it filled.
Changing your infusion set
Removing the reservoir
Each time you remove and replace a reservoir in your pump, you have to rewind your pump, and fill
the infusion set with insulin.
1 Remove the entire infusion set from your body.
2 If attached, remove the activity guard.
3 Turn the tubing connector half-turn counter clockwise, then pull the reservoir and connector out from
the pump.
tubing connector
half-turn counter-clockwise
3
4 Safely dispose of the used reservoir and infusion set item in a sharps container.
5 You must now rewind your pump as described in the next section.
Starting on insulin54
Przeglądanie stron 74