Delivery accuracy
• Testing has confirmed that the delivery accuracy for the maximum/minimum bolus/basal amounts are
within the +/-5 percent of the programmed value.
• All normal boluses will be delivered within a maximum of six minutes.
• Displacement (piston) accuracy: +/-2 percent.
Percent Error (%)
Delivery Volume Accuracy Curve (-1.20%) Overall Error
(MMT-523 Model) QTP6303 Qualification Pump (0.01 ml/hr 10/22/07 S/N: PAR162804U); 23°C,
MMT-326A Reservoir Lot#: WE0107318, MMT-317 Infusion set Lot: 620699
Time Interval (Minutes)
Maximum Error (percent)
Minimum Error (percent)
Overall Flow Error (percent)
-20
-15
-10
-5
0
5
10
15
20
0 153045607590105
Pump motor
The pump motor has a unique, patented design with an integrated safety check system. The system
delivers in precise increments.
Pump specifications180