Insulin delivery default settings
Menu Item Default
Setting
Limits Increments
Bolus Menu: Bolus Wizard feature: Off
Easy bolus: Off
Easy bolus step: 0.1 u 2.0 u 0.1 u
Scroll Rate: 0.10 u 0.025 u; 0.05 u; 0.10 u
Dual/Square bolus: Off
Maximum bolus: 10.0 u 0.0–25.0 u (per
single bolus)
BG Reminder: Off 0:00 - 5:00 0:30
Basal Menu: Patterns: Off
Maximum basal rate: 2.00 u/h 0.000–35.0 u/h 0.025 u (for 0.025–
0.975 u/h); 0.05 u (for
1.00–9.95 u/h); 0.1 u (for
rates of 10 u/h or more)
Basal rate: 0.000 u/h 0.025 u (for 0.025–
0.975 u/h); 0.05 u (for
1.00–9.95 u/h); 0.1 u (for
rates of 10 u/h or more)
Temp basal type: U/H Max basal rate
setting
0.025 u (for 0.025 -
0.975 u/h); 0.05 u (for
1.00 - 9.95 u/h);0.1 u (for
rates of 10 u/h or more);
(or 1%)
Utilities
Menu:
Lock Keypad: Off
(Alarm) History: (no defaults)
Alert type: audio, beep-
med
Auto-off: Off
Pump specifications192