Recommended separation distances between portable and mobile RF communications
equipment and the Paradigm 523/723/523K/723K pump system
This section provides information on the recommended separation distance between portable and
mobile RF communications equipment and the Paradigm 523/723/523K/723K pump. The Paradigm
523/723/523K/723K pump is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. Paradigm 523/723/523K/723K pump users can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Paradigm 523/723/523K/723K pump as
recommended below, according to the maximum output power of the communications equipment.
Separation distance according to the frequency of transmitter (m)
Rated maximum
output power of
transmitter (W)
80MHz to 800MHz
d = 0.35
P
800MHz to 6.0GHz
d = 0.70
P
0.01 0.035 0.07
0.1 0.11 0.11
1 0.35 0.7
10 1.1 2.2
100 3.5 7
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Pump specifications 203
Chapter 15