Example 1: Blood glucose on target (normal blood glucose) and no active insulin
Michael awakens in the morning before school and his mother has breakfast waiting for him. Before he
begins eating, he tests his blood glucose with his meter and his blood glucose result of 6.6 mmol/L (120
mg/dL) is automatically sent to his pump.
He estimates that his meal consists of 60 grams of carbohydrates. When prompted by the Bolus W
izard
feature, he enters this amount in the ENTER FOOD screen. Based on his Bolus Wizard feature settings,
the pump will suggest that he take 4.0 units of insulin.
(food estimate)
60 g
15 g/u
= 4 units
+
(correction estimate)
= 4 + 0
= 4 units
bolus estimate
Correction is 0 because the
current blood
glucose reading
is within the BG Target Range.
Example 2: Blood glucose above target (high blood glucose) and no active insulin
The next day, Michael wakes up before school. Before eating the same breakfast, he tests his blood
glucose with his meter and finds it to be 200 mg/dL (11.1 mmol/L), which is above his target of 120
mg/dL (6.6 mmol/L). His blood glucose reading is automatically sent to his pump.
When prompted by the Bolus Wizard feature, he enters his carbohydrate amount of 60 grams in the
ENTER FOOD screen. Based on his settings, the pump will suggest that he take 6.0 units of insulin.
(food estimate)
60 g
15 g/u
= 4 units
= 2 units
+
(correction estimate)
= 4 + 2
= 6 units
bolus estimate
11.1 mmol/L - 6.6 mmol/L
2.2 mmol/L/u
= 2 units
200 mg/dL - 120 mg/dL
40 mg/dL/u
Using the Bolus Wizard feature 87