My pump will not display my blood glucose reading from my meter
1 Make sure you are using the correct meter (a blood glucose meter supported by MWT1 technology).
Y
our pump will communicate with this meter only.
2 Make sure your meter's RF (radio frequency) feature is on and working correctly.
3 Make sure the meter option in your pump is set to On and you have set the meter ID number correctly
on the pump.
4 Make sure your pump does not have a low battery alert condition.
5 Make sure the meter is within 1.2 meters (4 feet) of the pump without anything in between such as
another person, a wall, etc.
6 Make sure there is no RF (radio frequency) interference from other electronic devices that could
prevent communication. These devices can include some cell phones, cordless phones, televisions,
computers, radios, other Paradigm pumps, meters and pump remote controls. To restore
communication, simply move away from these other types of devices, or turn them off.
7 Your pump will not show another reading. Make sure the pump is idle and the HOME screen is blank.
8 If your pump still does not receive your blood glucose reading from the meter, use the up/down
buttons to manually enter your blood glucose (in the ENTER BG screen).
I dropped my pump
Take care to protect your pump from being dropped.
1 Check that all connections are still tightly in place.
If you notice that the drive support cap is loose, or sticking out, discontinue use of the pump. Do not
press on the drive support cap. Pushing on the drive support cap may result in unintended insulin
delivery, which can cause hypoglycemia. See Warnings in the Introduction chapter for more
information.
2 Check the LCD, keypad and pump case for cracks or damage.
3 Check infusion set, including the tubing connector and tubing for cracks or damage.
4 Review the status screen, basal rates and other pump settings.
5 Perform the Selftest procedure located in the UTILITIES MENU.
6 Call your local Medtronic Diabetes representative for assistance.