Clearing the settings
T
ake the following steps only if you want to clear your pump to factory default settings.
WARNING: Do not clear your pump settings while it is connected to your body.
CAUTION: Do not clear your pump settings unless directed by your healthcare professional or a
Medtronic Diabetes representative. If you clear your pump settings, it will be necessary to reprogr
am
all your personal pump settings as directed by your healthcare professional. Additionally, you will
have to rewind your pump.
1 Go to the UTILITIES MENU screen and select User Settings.
Main > Utilities > User Settings
2
Hold down and press ACT.
3 The USER SETTINGS screen is displayed. Select Clear Settings, then press ACT.
4 A CONFIRM message displays, giving you the option of clearing your pump settings or not. Select YES
and press ACT to clear the settings. Y
ou can press ESC if you want to cancel.
5 You will see a RESET screen, and then the pump will go through various screens while it restarts. After
the pump clears all of your settings, the screen will go to the TIME/DATE SETUP screen.
6 Reset the time and date as described in Setting the time and date section in the Basic programming
chapter.
7 After you set the time and date, you must rewind your pump. Refer to the Rewinding your pump
section in the Starting on insulin chapter for instructions. Remember, all your settings have been
cleared and you must either restore or reprogram all your settings.
History
If you want to v
iew a listing of the dates and times of all recent User Settings operations you have
done on your pump, such as saves and restores, take the following steps:
1 Go to the UTILITIES MENU screen and select User Settings.
Main > Utilities > User Settings
2
Hold down
and press ACT.
3 The USER SETTINGS screen is displayed. Select History, then press ACT.