Filling the reservoir
Warning: Do not use the reservoir or infusion set if any liquid gets on the top of the reservoir or inside
the tubing connector (as shown in the image). Liquid can tempor
arily block the vents. This may result
in the delivery of too little or too much insulin, which can cause hypoglycemia or hyperglycemia. If
any liquid gets on the top of the reservoir or inside the tubing connector, start over with a new
reservoir and infusion set.
Tubing
connector
Top of
reservoir
WARNING: Using cold insulin can cause air bubbles in the reservoir and tubing. If your insulin is stored
in the refriger
ator, allow it to reach room temperature before filling your reservoir. When filling the
reservoir, take care to remove air bubbles.
CAUTION: Insulin comes in different vial types, depending on your country. If your vial is different
than the one described in the User Guide, consult with your healthcare professional on how to fill
your reservoir
.
1 Remove reservoir from package. Make sure plunger rod is fully extended.
2 Swab vial with alcohol (not shown).
3 Making sure you do not push down on the plunger, press the transfer guard onto the vial.
Starting on insulin58