2 Select the basal pattern you want to program, then press ACT.
3 The SET BASAL RATE 1 screen appears. The basal rate flashes, indicating that it can be changed. Set
your first rate, then press ACT.
The start time for your first basal rate is midnight and cannot be changed.
4 The SET START TIME 2 screen appears. The dashes under the screen name flash. The first basal rate is
now set.
If you want to use the same rate for the whole day
, press ESC and skip to step 10. If you want to
program another basal rate, follow steps 5 through 8.
5 In the SET START TIME 2 screen enter the time of the day you want this basal rate to become active.
6 Press ACT. The SET BASAL RATE 2 screen appears.
7 The previously set basal rate or dashes flash on the screen. Select the value for this basal rate.
8 Press ACT. The SET START TIME 3 screen appears. The second basal rate is now set.
9 If you do not need to set up any more basal rates, press ESC. If you need to set up more basal rates,
repeat steps 5 through 8 above for each rate.
10 After you press ESC, the BASAL RATE screen appears. The screen will show:
• the current basal pattern and basal rate,
• time it started, and
• the 24-hour basal total.
After you make changes to a pattern, the pump will use that pattern as the current basal. Make sure
the basal you want is selected in the SELECT PATTERN screen.
Select a pattern
Before you select a pattern to be active, make sure the P
atterns feature is turned on. After your standard
pattern and/or pattern A or B are set, do these steps to select a pattern to be the active one:
1 Go to the SELECT PATTERN screen.
Main > Basal > Select P
atterns
2 Select the desired pattern, then press ACT.
3 The screen will return to the BASAL MENU. Your basal pattern is now active. Exit the menus.
NOTE: If pattern A or B is active, the pump is in Special mode. An open circle appears at the top of the