METER BG NOW
Reason: A meter BG measurement is needed immediately to calibrate your sensor so that you can
continue receiving sensor glucose readings.
Action: Take and enter a meter blood glucose measurement. Follow the guidelines in the Calibrating the
sensor section in the Sensor features chapter.
METER BG BY
Reason: A meter BG measurement must be entered by the time that is shown to calibrate your sensor. The
METER BG BY alert is also known as the CAL REMINDER alert.
Action: Enter a blood glucose measurement to avoid a METER BG NOW alert.
LOW XX MG/DL (XX = SG measurement)
Reason: The sensor glucose number is equal to or lower than your low glucose limit. If you do not set a
Low Glucose Limit and do not set the glucose alerts ON, then you will not get a Low sensor glucose alert.
The pump plays four consecutive tones, in falling pitch, if an audible beep has been selected as the alert
type.
HIGH XXX MG/DL (XXX = SG measurement)
Reason: The sensor glucose number is higher than or equal to your high glucose limit. If you do not set a
High Glucose Limit and do not set the glucose alerts ON, then you will not get a High sensor glucose alert.
The pump plays four consecutive tones, in rising pitch, if an audible beep has been selected as the alert
type.
SENSOR ERROR
Reason: The sensor signals are either too high or too low.
Action: You do not need to change the sensor. Clear the alert. If the alert persists, test your transmitter
with the tester.