2 Connect the transmitter to the charger by lining it up, flat side down, with
the charger
. Push the two components together fully. Always allow at least
one minute before disconnecting the transmitter from the charger or the
transmitter may not work properly. If you disconnect the transmitter
before one minute, reconnect it to the charger for at least one minute.
3 Within 10 seconds after the transmitter is connected, a green light on the charger will flash for one to
two seconds as the charger powers on. For the rest of the charging time, the charger's green light will
flash in a continuing pattern of four flashes, pause, four flashes, pause.
green light
4 When charging is complete, the green light on the charger will stay on, without flashing, for 15 to
20 seconds and then turn off.
5 After the green charger light turns off, disconnect the transmitter from
the charger
. The green light on the transmitter will flash for approximately
five seconds and then turn off.
6 If the green light on the transmitter does not flash, reconnect it to the charger for at least one
minute.
7 After removing the transmitter from the charger, wait at least one minute before connecting it to a
sensor or tester
.
Sensor features118