Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 522/722/522K/722K pump system (consisting of the MMT-522/722/522K/722K pump
and MMT
-7703 transmitter) is intended for use in the electromagnetic environment specified below.
The customer or the user of the Paradigm 522/722/522K/722K pump system should make sure that
it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 522/522K/722/722K pump system is intended for use in the electromagnetic
env
ironment specified below. The customer or user of the Paradigm 522/522K/722/722K pump
system should make sure that it is used in such an environment.
Immunity
Test IEC 60601
Level
Compliance
Level Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
10Vrms Not applicable
Portable and mobile RF communications equipment
should be used no closer to any part of the
Paradigm 522/722/522K/722K pump, including
cables, than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Not applicable
Pump specifications200
Przeglądanie stron 218