BAD TRANSMTR
Reason: The transmitter battery is depleted.
Action: Recharge your transmitter immediately. See the Charging the transmitter section in the Sensor
features chapter.
BAD SENSOR
Reason: You may receive this alert after you receive two Cal Errors in a row, without the Cal Errors, or
when initializing your sensor.
Action: If the alert resulted from two Cal Errors in a row, replace the sensor. If the alert happened without
two Cal Errors, use the transmitter's tester to make sure your transmitter is working properly. If you
received this alert during initialization, you may be able to resolve this alert without replacing your
sensor. Call the HelpLine for further assistance.
SENSOR END
Reason: The sensor has reached the end of its life.
Action: Replace your sensor. The sensor has a maximum life of 72 hours (3 days). The 72-hour life span of
the sensor begins when the pump receives the first METER BG NOW alert.
CAL ERROR
Reason: An error occurred when entering a new meter BG measurement to calibrate the system. Some
possible causes are:
An incorrect blood glucose number was entered from the meter into the pump.
The entered blood glucose measurement was not current.
Your blood glucose is rising or falling rapidly.
The sensor needs more time to stabilize after being inserted.
The sensor is no longer reading the sensor glucose correctly.
Action: If you receive a Cal Error, wait until your BG is stable to enter a new meter BG for calibration. If
you receive a Cal Error on your second calibration, a BAD SENSOR alert will occur. Call the HelpLine if you
have questions.
Troubleshooting and alarms162