8 Press ACT to accept and deliver the bolus. The
BOLUS DELIVERY screen appears with an open circle
indicating that your pump is in Special Mode. The pump beeps or vibrates at the start of the bolus.
During bolus delivery, the pump returns to the HOME screen. If you want to see the progress of the
delivery, press ESC to see the STATUS screen. The pump beeps or vibrates at the end of the bolus, and
the open circle disappears.
Easy bolus
The EASY BOLUS button
allows a quick way to deliver a Normal bolus. You will pre-set the settings for
this feature in the
EASY BOLUS OPTION screen in the BOLUS MENU. Your pump is sent from the factory with
the Easy Bolus feature set to on. If you do not want to use Easy Bolus, turn it off.
After you set up Easy Bolus, with each
press, you can increase the Normal bolus amount by a fixed
amount, called a step. Before you can deliver an Easy Bolus, you must set the amount in the
EASY BOLUS
ENTRY screen. This amount equals the number of units of insulin for each step. The maximum number of
steps can equal up to your maximum bolus limit. When using vibrate mode, EASY BOLUS is limited to 20
steps or maximum bolus, whichever comes first.
After you set your step amount, you can program your Easy Bolus. When you are in the HOME screen, each
time you press
the Easy Bolus amount increases by one step. You will hear a beep or feel a vibration
for each step increase. Each beep is a different tone. This makes it easy for you to count the beeps while
you are programming your Easy Bolus.
Easy bolus setup
1 Go to the EASY BOLUS OPTION screen.
Main > Bolus > Easy Bolus
Select On/Set, then press ACT. If you do not want to use Easy Bolus, select Off and press ACT.
NOTE: If you are using the remote control, the Easy Bolus must be set to on.
Step value setup
Y
ou can set the step value from 0.1 to 2.0 units, or to the set maximum bolus limit of less than 2.0 units
(factory setting: 0.1). Set the step to a number that is convenient to use and easy to multiply.
Optimizing pump therapy 97
Chapter 7