Performance results in children and adolescents
The performance of the Guardian R
T in Children and Adolescents was evaluated in a clinical study.
Guardian RT results were compared to capillary blood glucose measured by a LifeScan OneTouch
®
Ultra
®
Meter. Sixty subjects with Type I diabetes participated in a three-site out-patient study. Subjects ranged in
age from 7 to 17 years old. Each subject wore a Guardian RT system, which records a sensor glucose value
every five minutes. Subjects were instructed to perform at least seven blood glucose meter measurements
each day.
Accuracy of Guardian RT readings
In this study, each blood glucose meter measurement was paired with the corresponding Guardian RT
reading (taken every 5 minutes). Pairing was done by selecting the Guardian RT value closest in time to
the blood glucose meter result. Agreement was analyzed by comparing paired glucose measurements.
Agreement between the matched pair was estimated by evaluating the difference between the Guardian
RT reading and the blood glucose meter measurement. The difference between them was calculated as a
percentage of the blood glucose meter measurement (Mean Absolute Percent Difference). The bias was
also calculated, and it is defined as the overall difference between the Guardian RT glucose values and the
blood glucose meter values. The paired glucose measurements are summarized in the following table
Number of Paired Glucose Measurements 2599
Mean Absolute Percent Difference (± SD) 19.0 ± 19.7%
Bias -6.0 mg/dL (-0.3 mmol/L)
Sensor accuracy 223
Appendix A