Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 522/722/522K/722K pump system (consisting of the MMT-522/722/522K/722K pump
and MMT
-7703 transmitter) is intended for use in the electromagnetic environment specified below.
The customer or the user of the Paradigm 522/722/522K/722K pump system should make sure that
it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Not applicable
±8 kV air
±30 kV air
(<5% relative
humidity)
The Paradigm
522/722/522K/722K pump
system should not be affected
by electrostatic discharge that
might occur under normal
conditions of use.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable
Not applicable
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not applicable
Not applicable
Not applicable
Not applicable
Pump specifications198