Manual prime
Manual prime fills the infusion set tubing with insulin before you insert the set into the body
. Manual prime
is only available after you rewind your pump.
WARNING: Make sure that the infusion set is disconnected from your body before you press ACT to
prime the pump. Never insert the reservoir into the pump while the tubing is connected to your body
.
Doing so could result in an accidental infusion of insulin.
1 After you rewind your pump, the PREPARING TO PRIME screen appears.
The pump may automatically return to the
HOME screen while you install the reservoir. Press ACT to
return to the PREPARING TO PRIME screen.
2 Press and hold ACT to start the prime. The pump will beep six times when the prime starts.
3 While you hold ACT, the pump will again beep six times when the screen begins counting the prime
units being used.
4 Continue to hold ACT until insulin droplets form on the tip of the infusion set needle, then release.
Make sure there are no bubbles in the tubing.
WARNING: If the PRIMING-HOLD ACT screen does not display, do NOT insert the infusion set in
your body. Please contact your local representative for assistance.
If your manual prime uses more than 30U insulin, the pump will show the following screen: IS PRIMING
COMPLETE?
When you get this message, do the following:
a. Make sure you are not connected to the pump.
b. Read the message on the screen, then press ESC, ACT to clear.
c. If your manual prime is complete (you see droplets on the tip of the infusion set needle), press
ESC and continue to step 5.
5 Press ESC. Your manual prime is complete.
6 You can now insert the infusion set into your body as described in the next section.
Starting on insulin64