2 Select Selftest, then press ACT.
3 Periodically, you will hear beeps as different mechanisms in the pump are being tested. As part of the
selftest, the pump will do these tests:
a. Screen T
est:
The screen will appear all black.
b. Selftest:
The pump will count down from 10.
c. Tone Test:
You should hear beeps.
d. Vibrate Test:
You will feel vibrations.
4 After the selftest is finished, TEST COMPLETE screen appears. The screen will return to the UTILITIES
MENU, then to the HOME screen.
User settings
The user settings function allows you to save, restore, and clear all pump settings. Y
ou can also view a
listing of the dates and times of all recent user settings operations you have done. The Save Settings
feature lets you keep a set of pump settings that you can restore to your pump if it is cleared or you need
to go back to these settings for any reason.
When you clear your pump, the pump settings are restored to the factory defaults, and you must either
use Restore Settings, if you have saved a set of pump settings, or reprogram all your settings before you
can use your pump again. The pump does not clear the internal pump memory.
WARNING: Do not clear your pump settings while it is connected to your body.
CAUTION: Do not clear your pump settings unless directed by your healthcare professional or a
Medtronic Diabetes representative. If you clear your pump settings, it will be necessary to reprogr
am
all your personal pump settings as directed by your healthcare professional. Additionally, you will
have to rewind your pump.
Saving the settings
Do these steps to save your current pump settings:
1 Go to the UTILITIES MENU screen and select User Settings.
Utilities 145
Chapter 10