Medtronic Paradigm 522 Podręcznik Użytkownika Strona 218

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 217
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 522/722/522K/722K pump system (consisting of the MMT-522/722/522K/722K pump
and MMT
-7703 transmitter) is intended for use in the electromagnetic environment specified below.
The customer or the user of the Paradigm 522/722/522K/722K pump system should make sure that
it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycle
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Pump specifications 199
Chapter 15
Przeglądanie stron 217
1 2 ... 213 214 215 216 217 218 219 220 221 222 223 ... 261 262

Komentarze do niniejszej Instrukcji

Brak uwag