plunger
rod
transfer
guard
reservoir
vial
31
4 Push down on the plunger to pressurize the vial. Hold down the plunger rod.
5 While still holding the plunger rod, flip the vial over so the vial is on top, slowly pull down on the
plunger to fill the reservoir.
6 Gently tap the side of the reservoir to make any air bubbles rise to the top of the reservoir.
7 Slowly push up on the plunger just enough to remove any air bubbles from the reservoir.
8 Slowly pull down on the plunger to fill the reservoir to the number of units desired.
9 To avoid getting liquid on the top of the reservoir, flip the vial over so that it is upright. Turn the
reservoir counter-clockwise, then pull straight up to remove the reservoir from the transfer guard.
10 Place the tubing connector onto the reservoir. Turn the connector clockwise, pressing gently against
the reservoir until you feel it slide in. Push in and continue turning until the reservoir and the
connector lock with a click.
Starting on insulin 59