Guidance and manufacturer's declaration
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Paradigm 522/722/522K/722K pump system (consisting of the MMT-522/722/522K/722K pump
and MMT
-7703 transmitter) is intended for use in the electromagnetic environment specified below.
The customer or the user of the Paradigm 522/722/522K/722K pump system should make sure that
it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The Paradigm 522/722/522K/722K pump system
uses RF energy only for system communication
functions. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Class B
Not applicable
Not applicable
The Paradigm 522/722/522K/722K pump system is
suitable for use in all establishments other than
domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
NOTE: The preceding statement is required by
IEC 60601-1-2 for Group 1, Class B devices.
However
, since the Paradigm 522/722/522K/722K
pump system is battery powered, its emissions
will not be affected by the establishment power
supply and there is no evidence of any issues
associated with the use of the system in domestic
establishments.