Basic programming
Setting the time and date
Setting the correct time and date in your pump is necessary for accurate basal insulin delivery and allows
you to keep an accurate record of your insulin delivery and other pump functions. Y
ou can select a 12-hour
or 24-hour clock. You must reset the time and date if you receive a CHECK SETTINGS alarm or you clear
your settings (Clear Settings function).
1 Go to the TIME/DATE
SETUP
screen.
Main > Utilities >
Time/Date
ACT
2 Select 12-Hour Setup or 24-Hour Setup and press ACT.
3 Press ACT again to change the settings.
4 Change each of the settings as follows:
a. Hour
Change the hour. Press ACT.
For 12-hour setups, press or until the correct A (am)
or P (pm) appears.
b. Minutes
Change the minutes. P
ress ACT.
Basic programming 39