The transmitter
The Medtronic MiniLink T
ransmitter (MMT-7703) is a device that takes electronic signals generated by the
glucose sensor and sends them by radio frequency to the pump. It has a tester (MMT-7706) and a charger
(MMT-7705).
MiniLink
T
ransmitter
MiniLink
Tester
MiniLink
Charger
Transmitter charger
The transmitter contains a non-replaceable, rechargeable battery that you can recharge as needed with
the charger
. The charger has a green light that shows the charging status and a red light that
communicates any problems during charging. If you see a red light, see the Understanding your
transmitter, tester, and charger section in the Troubleshooting and alarms chapter. The charger needs a
AAA or LR-03 battery to operate. A new AAA battery contains enough power to recharge the transmitter
more than 40 times.
Charging the transmitter
Before using the transmitter for the first time, you must fully charge the transmitter battery, which may
take up to eight hours. It is recommended to recharge the transmitter after each sensor use. If you choose
to recharge the transmitter after each sensor use, the charging time will be less than 20 minutes. A fully
charged transmitter battery will work more than 14 days without recharging. After 14 days of use, the
transmitter will fully recharge in less than two hours.
1 If a green light on the transmitter is lit or flashing, do not connect it to the charger. The transmitter
will not charge with its green light on. W
ait for the green light to turn off, then connect the
transmitter to the charger.