Setting the Cal Reminder
The Cal R
eminder feature allows you to set a reminder to calibrate your system. For example, if you set
your reminder to four hours, then you will receive a METER BG BY alert (Cal Reminder) four hours before
the next meter blood glucose (BG) entry is due (eight hours after your last successful sensor calibration).
To set up the Cal Reminder:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2 Select Cal Reminder, then press ACT. The
CAL REMINDER screen shows On.
3 Press ACT. The SET CAL REMINDER screen flashes the Cal Reminder default time of 1 hour (1:00).
4 Select your Cal Reminder time between 5 minutes (0:05) and 4 hours (4:00).
5 Press ACT. The EDIT SETTINGS screen shows the Cal Reminder time you selected. You are now ready to
enter the ID number of your transmitter.
Entering the transmitter ID
The transmitter ID (serial number) starts with 2 and is found on the flat side of your transmitter
. You must
enter the transmitter ID so that the transmitter and the pump can communicate with each other.
Write the Transmitter ID of the transmitter you are using here: __________________.
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2 Select T
ransmtr ID, then press ACT. The SET TRANSMITTER ID screen appears. The screen shows seven
dashes for the Transmitter ID, the first of which (on the left) is flashing.
3 Select the first number of the Transmitter ID (the number on the screen must
match the first number of the Transmitter ID). The transmitter ID is located on
the transmitter label, as shown.
4 Press ACT. The second dash flashes.
5 Select the next number of the Transmitter ID, then press ACT.
6 Repeat step 5 until all seven digits of the Transmitter ID have been selected.
Sensor features 115