BUTTON ERROR
Occurs if a button has been continually pressed for more than three minutes.
CHECK SETTINGS
When this alarm is active, you should check and/or reprogram your pump settings, including the time/
date.
E (ERROR)
An E-error alarm will show the letter E followed by two numbers. E-alarms cause all insulin delivery to
stop, the pump resets, and all your settings are cleared. After receiving this alarm, note the error number
then call your local Medtronic Diabetes representative for assistance.
EMPTY RESERVOIR
There is no insulin in the reservoir. Change the reservoir immediately.
FAILED BATT TEST
The pump tests the voltage of each battery installed. This test ensures a battery with low voltage is not
used. If the battery does not have enough voltage, this alarm will occur. The pump will not function and
the battery must be replaced. (Always make sure that you install a NEW battery into the pump.)
MAX DELIVERY
This alarm alerts you when you have taken more insulin than expected based on Maximum bolus and
Maximum basal rates.
MOTOR ERROR
Insulin delivery has stopped. This alarm occurs if your pump detects a motor error.
Troubleshooting and alarms166